Ansur plug-in module to control the QA-90 Mk II Electrical Safety Analyzer.
The QA-90 Test Sequence has a parameter called "Power off delay time" which can be used to add a delay at the end of the sequence. When using this feature, you will need to add 6 seconds to any delay you specify. This is because the QA-90 incorrectly processes the value being provided for this delay. For example, if a 2 second delay is desired, use 8 seconds instead. If 5 seconds of delay is required, enter 11 as the value. Using a delay of 6 or less will result in no delay happening at all.